Hugh A. Klein

An ion trap design for a space-deployable strontium-ion optical clock

Alessio Spampinato, Jonathan Stacey, Sean Mulholland, Billy I. Robertson, Hugh A. Klein, Guilong Huang, Geoffrey P. Barwood, Patrick Gill

Abstract

Optical atomic clocks demonstrate a better stability and lower systematic uncertainty than the highest performance microwave atomic clocks. However, the best performing optical clocks have a large footprint in a laboratory environment and require specialist skills to maintain continuous operation. Growing and evolving needs across several sectors are increasing the demand for compact robust and portable devices at this capability level. In this paper we discuss the design of a physics package for a compact laser-cooled 88Sr+ optical clock that would, with further development, be suitable for space deployment. We review the design parameters to target a relative frequency uncertainty at the low parts in 10^18 with this system. We then explain the results of finite element modelling to simulate the response of the ion trap and vacuum chamber to vibration, shock and thermal conditions expected during launch and space deployment. Additionally, an electrostatic model has been developed to investigate the relationship between the ion trap geometrical tolerances and the trapping efficiency. We present the results from these analyses that have led to the design of a more robust prototype ready for experimental testing.

Analysis of thermal radiation in ion traps for optical frequency standards

Miroslav Doležal, Petr Balling, Peter B R Nisbet-Jones, Steven A King, Jonathan M Jones, Hugh A Klein, Patrick Gill, Thomas Lindvall, Anders E Wallin, Mikko Merimaa, Christian Tamm, Christian Sanner, Nils Huntemann, Nils Scharnhorst, Ian D Leroux, Piet O Schmidt, Tobias Burgermeister, Tanja E Mehlstäubler, Ekkehard Peik

Abstract

In many of the high-precision optical frequency standards with trapped atoms or ions that are under development to date, the AC Stark shift induced by thermal radiation leads to a major contribution to the systematic uncertainty. We present an analysis of the inhomogeneous thermal environment experienced by ions in various types of ion traps. Finite element models which allow the determination of the temperature of the trap structure and the temperature of the radiation were developed for 5 ion trap designs, including operational traps at PTB and NPL and further optimized designs. Models were refined based on comparison with infrared camera measurement until an agreement of better than 10% of the measured temperature rise at critical test points was reached. The effective temperature rises of the radiation seen by the ion range from 0.8 K to 2.1 K at standard working conditions. The corresponding fractional frequency shift uncertainties resulting from the uncertainty in temperature are in the 10-18 range for optical clocks based on the Sr+ and Yb+ E2 transitions, and even lower for Yb+ E3, In+ and Al+. Issues critical for heating of the trap structure and its predictability were identified and design recommendations developed.